Warning: Undefined array key "options" in /home2/bptajhiz/domains/bptajhiz.com/public_html/wp-content/plugins/elementor-pro/modules/theme-builder/widgets/site-logo.php on line 192 AFIDA - Advanced Fuel Ignition Delay Analyser - BPtajhiz
, , ,

AFIDA – Advanced Fuel Ignition Delay Analyser

Company: Stanhope-Seta

Methods:ASTM D613, ASTM D8183, ASTM D975, BS EN 17155, BS EN ISO 5165 & BS 2000-41 & ISO 5165, IP 41, IP 617

Indicated Cetane Number Analysis

AFIDA is a revolutionary development providing you with a fully automated way of determining Indicated Cetane Number (ICN) of diesel and diesel related fuels. The AFIDA can analyse multiple samples in sequence and selects the samples using the 36-place autosampler. Only a small sample volume of 40 ml is required per test with a short analysis time of approximately 25 minutes.

The analyser uses a Constant Volume Combustion Chamber (CVCC) and incorporates a unique and patented high pressure injection system that generates fine fuel droplets similar to modern common rail injectors in most diesel engines. 

An ultra robust temperature controlled piezo electric injector provides rapid switching and highly repeatable fuel metering, offering you improved performance and consistency when compared with solenoid controlled injectors and pintle type nozzles.

 

Excellent correlation to ASTM D613, IP 41 and BS EN ISO 5165, BS 2000-41, ISO 5165
Wide sample types – middle distillates, blend components, biodiesel, FAME, XTL, HVO, heating oil and Jet A1
Suitable for use with Cetane improvers
Calibrated with industry approved PRFs over the range 35 to 85 ICN
Precise and reproducible software controlled determination of ICN
Results saved to PC or for download
Stable calibration – fixed test conditions eliminates routine temperature tuning
Low running costs, single gas supply for combustion air

 

Chamber temperature 580°C
Chamber pressure 17.5 bar
Injection pressure 1000 bar
Compressed air 20.9 ± 0.5 % O2
Sample volume approx. 40 ml for analysis and cleaning
Warm up time approx. 45 min
Analysis time approx. 25 min per sample

 

IP 617 Determination of generic cetane number (G-CN) of fuels using a constant volume combustion chamber: primary reference fuels calibration (PRFC) method
BS EN 17155 Liquid petroleum products – determination of indicated cetane number (ICN) of middle distillate fuels – primary reference fuels calibration method using a constant volume combustion chamber
ASTM D613 Standard Test Method for Cetane Number of Diesel Fuel Oil
IP 41 Determination of the ignition quality of diesel fuels — Centane engine method
BS EN ISO 5165 & BS 2000-41 & ISO 5165 Determination of the ignition quality of diesel fuels. Cetane engine method
ASTM D8183 Determination of Indicated Cetane Number (ICN) of Diesel fuel oils using a Constant Volume Combustion Chamber – reference fuels calibration method
ASTM D975 Standard Specification for Diesel Fuel Oils , biodiesel, biodiesel blend, diesel, fuel oil, petroleum and petroleum products

 

ASTM

, ,

BS EN

BS EN ISO

BS

ISO

IP

,

Standard Metod

, , , , ,

Scroll to Top